Introduction
HLA-E, HLA-F, and HLA-G are non-classical HLA class I genes that play a pivotal role in the precise regulation of immune responses, despite their limited polymorphism and tissue-specific expression. Consequently, genotyping these genes offers critical insights for clinical decision-making across various medical disciplines.
The AltruTypeTM Long-Read Sequencing HLA-E/F/G Genotyping Kit (Single-Molecule Sequencing Method) is engineered upon PacBio HiFi long-read sequencing technology. It delivers a specialized, high-resolution genotyping solution tailored specifically for the HLA-E, HLA-F, and HLA-G loci. By accurately resolving the sequence characteristics of these target genes, the kit provides robust, professional support for clinical applications in transplantation medicine, gynecology, and beyond, serving as an essential tool for achieving precision diagnosis and treatment.
This kit is intended for the identification and characterization of HLA complex genes from human genomic DNA isolated from whole blood. It targets HLA-E, HLA-F and HLA-G genes.
The workflow utilizes polymerase chain reaction (PCR) to amplify a predefined set of target genes. The resulting amplicons are designed for downstream library preparation and sequencing on PacBio HiFi platforms. The generated sequencing data is subsequently analyzed using Haorui TransMatchTM software to produce high-resolution HLA genotyping results.
Note: Test results obtained with this kit are for clinical reference only and should not be used as the sole basis for diagnosis or clinical decision-making. Clinicians should make comprehensive judgments and treatment decisions by integrating these results with the patient’s medical history, physical examination, and other laboratory findings.
The AltruTypeTM Long-Read Sequencing HLA-E/F/G Genotyping Kit (available in 24/48/96/192 reactions) is an integrated solution designed for the amplification and library preparation of HLA targets. The kit enables the simultaneous amplification of HLA-E, -F, -G, followed by streamlined library construction for sequencing.
The workflow proceeds as follows:
(1) Nucleic Acid Extraction: Genomic DNA is isolated from the sample.
(2) PCR Enrichment: Target regions are amplified via a single, long-range multiplex PCR covering all 3 loci.
(3) Library Preparation: PacBio’s proprietary SMRTbell adapters are ligated to the amplicons to construct circular DNA libraries.
(4) Sequencing: Libraries are loaded onto the Revio or Vega long-read sequencing platform using the HiFi sequencing mode.
(5) Signal Detection: Fluorescently labeled nucleotides are incorporated during polymerization, and real-time signals are captured to generate high-fidelity (HiFi) reads.
(6) Data Analysis: Reads are aligned to the human genome and HLA reference sequences (IPD-IMGT/HLA Database) using Haorui TransMatch software to export high-resolution HLA genotyping results.
Product Configuration
Reactions | Catalog |
24 | L1-120-130 |
48 | L1-120-120 |
96 | L1-120-110 |
192 | L1-120-100 |
Shipping and Storage
1. Storage Conditions
Store the kit at -20 ±5°C. The shelf life is 3 months from the date of manufacture (Note: Magnetic beads should be stored in a 4°C refrigerator).
2. Freeze-Thaw Stability
The kit may undergo no more than 6 freeze-thaw cycles. Once opened, use immediately.
3. Shipping Requirements
Transport under temperature-controlled conditions (-20 ±5°C).
Maximum allowable transit time: 3 days.
Upon receipt, verify that the internal temperature remains within -20 ±5°C before opening.
4. Refer to the lot-specific label for manufacturing date (MFG) and expiration date (EXP).